ISO 15632:2021 Microbeam analysis — Selected instrumental performance parameters for the specification and checking of energy-dispersive X-ray spectrometers (EDS) for use with a scanning electron microscope (SEM) or an electron probe microanalyser (EPMA)[

标准简介

Microbeam analysis — Selected instrumental performance parameters for the specification and checking of energy-dispersive X-ray spectrometers (EDS) for use with a scanning electron microscope (SEM) or an electron probe microanalyser (EPMA)是国际标准化组织(International Organization for Standardization,简称ISO)发布的ISO标准,适用于世界范围。

标准截图

Microbeam analysis — Selected instrumental performance parameters for the specification and checking of energy-dispersive X-ray spectrometers (EDS) for use with a scanning electron microscope (SEM) or an electron probe microanalyser (EPMA)
Microbeam analysis — Selected instrumental performance parameters for the specification and checking of energy-dispersive X-ray spectrometers (EDS) for use with a scanning electron microscope (SEM) or an electron probe microanalyser (EPMA)(截图)

 

标准文档说明

标准文档类型为Microbeam analysis — Selected instrumental performance parameters for the specification and checking of energy-dispersive X-ray spectrometers (EDS) for use with a scanning electron microscope (SEM) or an electron probe microanalyser (EPMA)高清PDF版本(文字版),标准文档内可进行搜索,可以复制原文,可粘贴。

标准部分原文

INTERNATIONAL STANDARD ISO 15632:2021

Microbeam analysis — Selected instrumental performance parameters for the specification and checking of energy-

dispersive X-ray spectrometers (EDS) for use with a scanning electron microscope (SEM) or an electron probe microanalyser (EPMA)

1 Scope

This document defines the most important quantities that characterize an energy-dispersive X-ray spectrometer consisting of a semiconductor detector, a pre-amplifier and a signal-processing unit as the essential parts. This document is only applicable to spectrometers with semiconductor detectors operating on the principle of solid-state ionization. This document specifies minimum requirements and how relevant instrumental performance parameters are to be checked for such spectrometers attached to a scanning electron microscope (SEM) or an electron probe microanalyser (EPMA). The [2] [3]

procedure used for the actual analysis is outlined in ISO 22309 and ASTM E1508 and is outside the scope of this document.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

ISO 23833, Microbeam analysis — Electron probe microanalysis (EPMA) — Vocabulary

ISO 22493, Microbeam analysis — Scanning electron microscopy — Vocabulary

3 Terms and definitions

For the purposes of this document, the terms and definitions given in ISO 23833, ISO 22493 and the following apply.

ISO and IEC maintain terminological databases for use in standardization at the following addresses:

— ISO Online browsing platform: available at https://www.iso.org/obp

— IEC Electropedia: available at http://www.electropedia.org/

NOTE With the exception of 3.1, 3.2, 3.2.1, 3.2.2, 3.9, 3.11, 3.12, 3.13 and 3.14, these definitions are given in [2] [4]

the same or analogous form in ISO 22309 , ISO 18115-1 and ISO 23833.

3.1

energy-dispersive X-ray spectrometer

device for determining X-ray signal intensity as a function of the energy of the radiation by recording the whole X-ray spectrum simultaneously

Note 1 to entry: The spectrometer consists of a solid-state detector, a preamplifier, and a pulse processor. The detector converts X-ray photon energy into electrical current pulses which are amplified by the preamplifier. The pulse processor then sorts the pulses by amplitude so as to form a histogram distribution of X-ray signal intensity (3.8) vs energy.

ISO 15632:2021

3.2

count rate

number of X-ray photons per second

3.2.1

input count rate

ICR

number of X-ray photons absorbed in the active detector area per second that are input to the electronics

3.2.2

output count rate

OCR

number of valid X-ray photon measurements per second that are output by the electronics and stored in memory, including sum peaks

Note 1 to entry: When the electronics measures individual X-ray photon energies, there is some dead time (3.4) associated with each individual measurement. Consequently, the number of successful measurements is less than the number of incident photons in every practical case. Thus, the accumulation rate into the spectrum (output count rate (3.2.2)) is less than the count rate of photons that cause signals in the detector (input count rate (3.2.1)). Output count rate (3.2.2) may be equal to input count rate (3.2.1), e.g. at very low count rates and for very short measurements.

3.3

real time

duration in seconds of an acquisition as it would be measured with a conventional clock

Note 1 to entry: For X-ray acquisition, in every practical case the real time (3.3) always exceeds the live time (3.5).

3.4

dead time

time during which the system is unable to record a photon measurement because it is busy processing a previous event (frequently expressed as a percentage of the real time (3.3))

Note 1 to entry: Dead time = real time – live time.

Note 2 to entry: Dead-time fraction = 1 − OCR/ICR.

3.5

live time

effective duration of an acquisition, in seconds, after accounting for the presence of dead time (3.4)

Note 1 to entry: Live time = real time for an analysis minus cumulative dead time.

Note 2 to entry: Live-time fraction = 1 − dead-time fraction.

3.6

spectral channel

discrete interval of the measured energy for the histogram of recorded measurements with a width defined by a regular energy increment

3.7

instrumental detection efficiency

ratio of quantity of detected photons and quantity of the photons incident on the active detector area

3.8

signal intensity

strength of the signal in counts per channel or counts per second per channel at the spectrometer output after pulse processing

Note 1 to entry: This definition permits intensity to be expressed as either “counts” or “counts per second” (CPS). The distinction is not relevant to the procedures described in this standard so long as either one or the other is consistently employed.